Model ES-215 ES-315
Chamber Capacity 22 liters 53 liters
Operating Temperature 105 to 123degree Celsius 105 to 132degree Celsius
Temperature control Microprocessor Control
Temperature Increments 1degree
Temperature Display Type Digital
Pressure Display Range 0 to 2.0 Kgf/sq. cm 0 to 3.5 Kgf/sq. cm
Pressure Display Type Analog
Maximum Operating Pressure 1.5Kgf/sq. cm 2.2Kgf/sq. cm
Sterilization Heat Source Stainless steel electric heater, 1.5Kw Stainless steel electric heater, 2.0Kw
Time Display 1 to 240minutes, and continuous(Sterilization Cycle)
1 to 8 hours(Warming Cycle)
Dimensions 400W x 460D x 920H mm 490W x 560D x 1,090H mm
Height of Worksurface 705mm 875mm
Chamber Diameter 248mm 325mm
Chamber Height 450mm 635mm
Net Weight 50Kg 80Kg
Power 110/120V,50/60Hz,15amps
220/230/240V,50/60Hz,7apms
110/120V,50/60Hz,20amps
220/230/240V,50/60Hz,10amps

Back 

 

Please send mail to Sales@Magenta-biomedika.com with questions or enquires about this product.