1. SD-320 autoclave equips drying feature is completely self-contained, compact and portable.
2. Top-loading configuration chamber has dimensions of 325mm diameter; 650mm depth and 55 liters capacity enable to sterilize various medical and laboratory materials.
3. Over pressure releasing safety valve, ground leakage breaker, water level sensor and overheat detector secure customer's safety.
4. If the temperature goes down during the sterilization, such as in case of slow air exhaust from the sample, the microprocessor of SD-320 detects temperature drops and starts energizing to the heater to get the correct temparure. During that procedure, the timer stops counting to avoid incorrect sterilization.

Next

 